Programming Module & Software
Temperature / Microprocessor Integrated Sensor Transmitter - MIST / MIST PKIT

Programming Module & Software Picture
  • MIST™ calibration software CD-ROM
  • Communication module with RS-232 or USB
  • 110 VAC Wall adapter (RS-232 only)

The MIST PKIT is used to re-calibrate and re-scale MIST Temperature Sensors™. Two methods of calibration are available. The first, called One Point Calibration, is performed by setting the MIST Temperature Sensor™ to a known temperature and adjusting the 4-20mA output to the corresponding value.

The second method, called Two Point Calibration, is performed by setting the MIST Temperature Sensor™ to a known point in the operating range and then recording this point by simply clicking the corresponding button. The next step is to set the MIST Temperature Sensor™ to a second known point in the operating range and record it in the same manner. Finally, by clicking on the “Process” button, the calibration parameters are automatically calculated and saved in the MIST Temperature Sensor™ memory.

The MIST PKIT software can also be used to re-scale the MIST Temperature Sensors™output within its operational limits. A choice of many pre-scaled values is offered, as well as a custom range. Once the range is selected, simply clicking on the “Set” button will automatically re-scale the output and save the new values in the MIST Temperature Sensor™ memory.

Programming Module & Software Details