Surface Mount Cylinder Pad Type
Temperature / Microprocessor Integrated Sensor Transmitter - MIST / MIST M344

Surface Mount Cylinder Pad Type Picture
  • Output signals: 4-20mA 2-wire; 0-5, 1-5, 0-10 VDC 3-wire
  • Input: Platinum RTD 100, Class A, 3-wire
  • Ranges: Std. and programmable from -50°C to 200°C
  • Power Supply: 10-32 VDC polarity protected
  • Accuracy: ±(.10°C + 0.10% of calibrated span ) with two-point calibration
  • Wetted Mat'l: SS316/316L, dual certified
  • Housing: Mat'l: SS316
  • Electrical Connection: M12 male receptacle
  • Ambient Temp. Range: -40°C to 80°C (housing only)
  • Protection: Meets NEMA 6P, IP67 (IEC 529)

The MIST M344 RTD Temperature Transmitter consists of surface mount cylinder pad type RTD, extension cable to remote transmitter and M12 connector. It is designed for heavy duty BTU temperature measurement applications due to its very high accuracy. As all the other MIST Series transducers, it can be re-calibrated and re-scaled in the field.

Surface Mount Cylinder Pad Type Details

  • Model Number

  • 1
    Calibrated Temperature Range

  • 2
    Output

  • 3
    Pad Diameter 'D'

  • 4
    Pad Material

  • 5
    Pad Length

  • 6
    Mounting Pad Radius “R”
      From 0 to 1 inch
  • 7
    Extension Cable Type

  • 8
    Extension Cable Length
      From 1 to 99 feet
  • 9
    Connector Type

  • 10
    Extension Cable Length “H”

  • 11
    Matched Pair Options


  • Comments / Specifications